Page 1 of Widex Bravo CIC Technical Guide by Hearing Planet
100 digital cic · programmable sp3 or compass · 2 channel dsp · wide dynamic range compression b2-cic is fine tuned with three parameters lf low frequency gain hf high frequency gain mpo maximum power output recommended for · mild to moderately severe hearing losses · all configurations of hearing loss including sloping flat and reverse slope losses · people wanting a fully automatic and discreet hearing aid suggested fitting range proper fitting of this cic hearing aid requires the taking of a deep-canal impression hearing healthcare professionals should not attempt this type of fitting unless they have the necessary skills to take this type of impression safely for further information please consult the senso cic impression manual the bravo digital signal processor dsp works with 20-bit representation of the signal and a sampling rate of 32 khz bravo b2-cic is a superbly engineered hearing aid in the renowned